112 Chapter 5
Disassembly
A2 Disassembly
A2 Disassembly
This section describes how to disassemble the A2 assembly. Use Table
5-1 below to see which assemblies are replaceable.
NOTE
Periodic Adjustment Intervals
The adjustment programs Periodic Calibration, IQ Calibration, and
Eb/No Calibration should be performed after any assembly referred to
in Table 7-1 on page 159 is replaced, or at least every 24 months. These
program can be run anytime to optimize the performance of the Test
Set. See Chapter 7 , “Periodic Adjustments,” on page 157 for details.
NOTE
Performance Test Intervals
The performance tests in Chapter 8 , “Performance Tests,” on page 173
should be performed when certain assemblies are repaired or replaced,
or at least every 24 months. See Table 3-2 on page 75 for those
assemblies requiring performance testing/calibration.
Module and PC Board Assemblies
1. Remove the Test Set’s external and top internal covers, see
“Removing the External and Internal Covers” on page 104.
2. Remove modules using a flat-blade screwdriver to pry them upward
from the chassis, see Figure 5-8 on page 113.
3. Remove PC board assemblies by lifting them from the chassis (some
assemblies have pull rings) and then disconnect any cables that may
be attached to it. See Figure 5-9 on page 114 for cable connections.
Table 5-1 A2 Assemblies
Module and PC Board Assemblies see page 112
Control Interface Assembly see page 117
PCMCIA Assembly see page 116
RF Input/Output, Upconverter, & Downconverter
Assemblies
see page 118
LO IF/IQ Modulator and CDMA Generator Reference (Gen
Ref) Assemblies
see page 120
see page 122
see page 124