Chapter 8 209
Performance Tests
Oscilloscope Amplitude Accuracy Performance Test 18
4. Set the frequency as shown in the Performance Test Record (PTR).
For each setting, perform the following:
a. Adjust the level until the digital multimeter reads 5 V.
b. Set Controls to Main and adjust the Time/Div on the Test Set to
display 2 to 3 cycles of the waveform.
c. Set Controls to Marker and press the knob (with the cursor in the
Marker To Peak+ field) to move the marker to the peak of the
waveform.
d. Read the Lvl and compare the reading to the limits in the PTR.