196 Chapter 8
Performance Tests
RF Generator Spurious Spectral Purity Performance Test 7
RF Generator Spurious Spectral Purity
Performance Test 7
The purpose of this test is to verify that the Test Set meets the
specification limits in PTR (Performance Test Record) Table 9-7, “RF
Generator Spurious Spectral Purity Test 7 Record,” on page 254.
Spurious signals with the carrier set to several frequencies and two
different levels (maximum output and minimum level vernier) are
searched for by an RF spectrum analyzer.
Figure 8-12 Setup
Procedure
1. Set up the spectrum analyzer in accordance with its operating
manual.
2. On the Test Set:
a. Press
Preset.
b. Select the INSTRUMENT CONFIGURE screen.
c. Set the RF Display field to Freq.
d. Select the CDMA GENERATOR screen.
e. Set the CW RF Path field to Bypass.
f. Select the RF GENERATOR screen.
g. Set the RF Gen Freq to 242 MHz.
h. Set the Amplitude to -10 dBm.
3. Set the test set’s RF generator to the frequencies and levels shown in
the Performance Test Record (PTR) and measure the level of the
spurious signals at the frequencies shown. For each measurement
convert the harmonic level to dB below the fundamental (dBc) and
compare the computed levels to the limits.
INPUT 50Ω
DUPLEX OUT
Spectrum Analyzer