204 Chapter 8
Performance Tests
AF Analyzer DC Level Accuracy Performance Test 15
AF Analyzer DC Level Accuracy
Performance Test 15
The purpose of this test is to verify that the Test Set meets the
specification limits in PTR (Performance Test Record) Table 9-15, “AF
Analyzer DC Level Accuracy Test 15 Record,” on page 265. To measure
DC level accuracy, a DC signal is measured by an external digital multi
meter and compared to the Test Set’s internal DC voltmeter reading.
Figure 8-20 Setup
Procedure
1. Set the multimeter to measure DC volts.
2. On the Test Set:
a. Press
Preset.
b. Select the AF ANALYZER screen.
c. Set the AF Anl In field to Audio In.
d. Set the Audio In Lo field to Gnd.
e. Select the RF GENERATOR screen.
f. Set the AFGen1 To field to Audio Out.
g. Set the AFGen1 Freq field to 0.0 Hz.
h. Set the Audio Out field to DC.
i. Set the level of Audio Frequency Generator 1 as shown in the
Performance Test Record and measure the DC level. Compare the
measured voltage to the limits.
INPUT
AUDIO OUT
Multimeter
AUDIO IN HI
Comentários a estes Manuais