160 Chapter 7
Periodic Adjustments
Periodic Adjustments
Equipment
Equipment for the Periodic Adjustments Programs
For the Timebase Reference Using a Counter calibration you will
need to connect a frequency counter to the rear-panel 10 MHz REF
OUTPUT connector. The accuracy of the counter will determine the
accuracy of the Test Set’s internal reference. You will use the counter
to set the timebase reference DACs.
For the Timebase Reference Using a Source calibration you will
need to connect a signal generator to the front-panel ANT IN
connector.
For the Voltmeter References calibration you will need a DC
voltmeter that can measure
±5 V with ±0.015% accuracy.
Figure 7-1 Periodic Adjustments Menu
percal.eps
SERVICE MENU
1
2
3
4
5
Help
Exit
Functional Diagnostics
AF Diagnostics
RF Diagnostics
CDMA Diagnostics
Edit RF Diagnostic Limits
Periodic Calibration
IQ Calibration
Eb/No Calibration
=>
Move pointer to desired program using the
knob then press the knob. Press Help for
information on the tests. Press Exit to abort.
SERVICE MENU Screen
1
2
3
4
5
Exit
Timebase Reference Using a Counter
Timebase Reference Using a Source
Voltmeter References
Audio Frequency Generator Gain
External Modulation Path Gain
Audio Analyzer 1 Offset
Variable Frequency Notch Filter
=>
Move the pointer to the desired test using the
knob then press the knob. Press Exit key to
abort.
Periodic
Calibration
Menu
Periodic Calibration Menu