Agilent 4294A PRECISION IMPEDANCE ANALYZER Service Manual
Manual Change
Agilent Part No. N/A
May 2009
Change 1
Change the Measurement Uncertainty value of Frequency Accuracy Test (Page 68) to the
following information and appended TAR to this report.
Frequency Accuracy Test
Without Option 1D5
Frequency Test Limit Test Result
1
Measurement
Uncertainty
TAR
10MHz ±200 Hz Hz ± 0.3 Hz 890
With Option 1D5
Frequency Test Limit Test Result
1
Measurement
Uncertainty
TAR
10MHz ±1.3 Hz Hz ± 0.2 Hz 890
Vista de página 0
1 2 3 4 5 6 ... 248 249

Resumo do Conteúdo

Página 1 - Manual Change

○C Copyright 2009 Agilent Technologies Agilent 4294A PRECISION IMPEDANCE ANALYZER Service Manual Manual Change Agilent Part No. N/A May 20

Página 2 - OSC Level Accuracy Test

2 NoticesThe information contained in this document is subject to change without notice.This document contains proprietary information that is prote

Página 3

92 3 AdjustmentHPOT-Amplifier Offset Correction ConstantsHPOT-Amplifier Offset Correction ConstantsThe purpose of this procedure is to adjust the HPO

Página 4

3 93AdjustmentRelative Impedance Correction Constants1. Chapter Title2. Chapter Title 3. Adjustment4. Chapter Title5. Chapter TitleRelative Impedance

Página 5

94 3 AdjustmentDC Bias Monitor Correction ConstantsDC Bias Monitor Correction ConstantsThe purpose of this procedure is to adjust the DC Bias Monitor

Página 6

3 95AdjustmentDC Bias Monitor Correction Constants1. Chapter Title2. Chapter Title 3. Adjustment4. Chapter Title5. Chapter TitleFigure 3-8Figure 3-8

Página 7

96 3 AdjustmentDC Bias Level Correction ConstantsDC Bias Level Correction ConstantsThe purpose of this procedure is to adjust the DC bias level.Requi

Página 8

3 97AdjustmentOSC Level Monitor Correction Constants1. Chapter Title2. Chapter Title 3. Adjustment4. Chapter Title5. Chapter TitleOSC Level Monitor C

Página 9 - Service Manual

98 3 AdjustmentOSC Level Monitor Correction ConstantsFigure 3-11 OSC Level Monitor Correction Constants Setup 14. Click OK, then wait till the measu

Página 10 - Manual Printing History

3 99AdjustmentOSC Level Correction Constants1. Chapter Title2. Chapter Title 3. Adjustment4. Chapter Title5. Chapter TitleOSC Level Correction Consta

Página 11 - Safety Summary

100 3 AdjustmentImpedance Measurement Correction ConstantsImpedance Measurement Correction ConstantsThe purpose of this procedure is to adjust the re

Página 12 - Certification

3 101AdjustmentImpedance Measurement Correction Constants1. Chapter Title2. Chapter Title 3. Adjustment4. Chapter Title5. Chapter TitleFigure 3-14 I

Página 13 - Assistance

3Safety SummaryThe following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Fail

Página 14 - Typeface Conventions

102 3 AdjustmentImpedance Measurement Correction ConstantsFigure 3-15 Impedance Measurement Correction Constants Setup 26. Follow the adjustment pro

Página 15 - Contents

3 103AdjustmentFrequency Reference Adjustment (Opt.1D5 only)1. Chapter Title2. Chapter Title 3. Adjustment4. Chapter Title5. Chapter TitleFrequency R

Página 16

104 3 AdjustmentFrequency Reference Adjustment (Opt.1D5 only)2. Adjust “TIME BASE ADJ” on the rear panel so that the counter reading is 10 MHz ± 0.1

Página 17

1054. Troubleshooting4 TroubleshootingThis chapter describes troubleshooting flow and provides the procedure to determine which group or assembly is

Página 18

106 4 TroubleshootingTROUBLESHOOTING SUMMARYTROUBLESHOOTING SUMMARYThe troubleshooting strategy of this manual is based on a verification (rather tha

Página 19

4 107TroubleshootingTROUBLESHOOTING SUMMARY4. TroubleshootingFigure 4-1 Troubleshooting Organization

Página 20

108 4 TroubleshootingSTART HERESTART HEREThis chapter contains the following troubleshooting procedures. By performing them in the order listed below

Página 21 - 1 General Information

4 109TroubleshootingINSPECT THE POWER ON SEQUENCE4. TroubleshootingINSPECT THE POWER ON SEQUENCECheck the FanTurn the analyzer power on. Inspect the

Página 22 - [System]

11 0 4 TroubleshootingINSPECT THE REAR PANEL FEATUREINSPECT THE REAR PANEL FEATUREIf the 4294A is operating unexpectedly after these checks are verif

Página 23

4 111TroubleshootingInternal Tests Failure Troubleshooting4. TroubleshootingInternal Tests Failure TroubleshootingCheck the Power-On SelftestThe anal

Página 24 - Require Equipment

4 Safety SymbolGeneral definitions of safety symbols used on the instrument or in manuals are listed below.Instruction Manual symbol: the product is

Página 25 - *3.furnished with the 4294A

Table 4-1 Troubleshooting Information for Internal Test FailureTest No.First Failed Test Probable Faulty AssemblyA1 A51 A50 A9 A2 A7 A3 A4 A5 A6 A81

Página 26 - General Information

4 113TroubleshootingExternal Tests Failure Troubleshooting4. TroubleshootingExternal Tests Failure TroubleshootingIf you can't identify the faul

Página 27 - 2 Performance Test

Table 4-2 Troubleshooting Information for External Test FailureTest No.First Failed Test Probable Faulty AssemblyA1 A51 A30 A53 A2 A7 A3 A4 A5 A6 A8

Página 28 - Performance Test

4 115TroubleshootingPerformance Tests Failure Troubleshooting4. TroubleshootingPerformance Tests Failure TroubleshootingPerform the following procedu

Página 29 - Test Equipment

Table 4-4 Troubleshooting Information for Performance Test Failure 2Test No.First Failed Test Probable Faulty AssemblyA1 A51 A50 A9 A2 A7 A3 A4 A5 A

Página 30

1175. Power SupplyTroubleshooting5 Power Supply TroubleshootingThis chapter describes troubleshooting procedures to isolate the faulty assembly with

Página 31

11 8 5 Power Supply TroubleshootingINTRODUCTIONINTRODUCTIONUse this procedure only if you have read Troubleshooting, and you believe the problem is i

Página 32

5 119Power Supply TroubleshootingINTRODUCTION5. Power SupplyTroubleshootingFigure 5-1 Power Supply Lines Simplified Block Diagram

Página 33

120 5 Power Supply TroubleshootingSTART HERESTART HERE1. Check the Fan Look at the fan on the rear panel. Check that the fan is rotating.• If the fan

Página 34 - [Center] - [1] - [M/μ]

5 121Power Supply TroubleshootingSTART HERE5. Power SupplyTroubleshootingA50 Shutdown LEDThe A50 SHUTDOWN LED turning off indicates some of A50 power

Página 35

5WarrantyThis Agilent Technologies instrument product is warranted against defects in material and workmanship for a period corresponding to the indi

Página 36

122 5 Power Supply TroubleshootingSTART HEREFigure 5-3 A1 +5 VD LED LocationMeasure the A1 +5 VD VoltageMeasure the DC voltage on a test point A1TP4

Página 37 - [Center] - [1] - [M/m]

5 123Power Supply TroubleshootingSTART HERE5. Power SupplyTroubleshootingFigure 5-4 A9 Two LEDs LocationDisconnect Wire from A3 P11Turn the analyzer

Página 38

124 5 Power Supply TroubleshootingSTART HEREFigure 5-5 A3 P11 and RF Cable “M” location

Página 39 - DC Bias Monitor Accuracy Test

5 125Power Supply TroubleshootingFIND OUT WHY THE FAN IS NOT ROTATING5. Power SupplyTroubleshootingFIND OUT WHY THE FAN IS NOT ROTATINGIf the fan is

Página 40

126 5 Power Supply TroubleshootingFIND OUT WHY THE FAN IS NOT ROTATINGFigure 5-7 A40 Output Voltage3. Check the A50 SHUTDOWN LEDWhen the fan stops,

Página 41

5 127Power Supply TroubleshootingFIND OUT WHY THE A50 SHUTDOWN LED IS OFF5. Power SupplyTroubleshootingFIND OUT WHY THE A50 SHUTDOWN LED IS OFFUse th

Página 42

128 5 Power Supply TroubleshootingFIND OUT WHY THE A1 +5 VD LED IS NOT ON STEADILYFIND OUT WHY THE A1 +5 VD LED IS NOT ON STEADILYIf the +5 VD LED is

Página 43 - DC Bias Level Accuracy Test

5 129Power Supply TroubleshootingFIND OUT WHY THE A1 +5 VD LED IS NOT ON STEADILY5. Power SupplyTroubleshootingFigure 5-8 A1 CPU Connector Location3

Página 44

130 5 Power Supply TroubleshootingFIND OUT WHY THE A1 +5 VD LED IS NOT ON STEADILY3. Reinstall the removed assemblies one at a time. Turn the analyze

Página 45

5 131Power Supply TroubleshootingTROUBLSHOOT THE FAN AND THE A50 DC-DC CONVERTER5. Power SupplyTroubleshootingTROUBLSHOOT THE FAN AND THE A50 DC-DC C

Página 46 - Measurement Accuracy Test

6 For any assistance, contact your nearest Agilent Technologies Sales and Service Office. Addresses are provided at the back of this manual.Typeface

Página 47

132 5 Power Supply TroubleshootingTROUBLSHOOT THE FAN AND THE A50 DC-DC CONVERTER2. Troubleshoot the A50 DC-DC Converter1. Turn the analyzer power of

Página 48

1336. Digital ControlTroubleshooting6 Digital Control TroubleshootingThis chapter describes troubleshooting procedures to isolate the faulty assembl

Página 49

134 6 Digital Control TroubleshootingINTRODUCTIONINTRODUCTIONUse this procedure only if you have followed the procedures in the Troubleshooting chapt

Página 50 - Capacitor) Setting

6 135Digital Control TroubleshootingINTRODUCTION6. Digital ControlTroubleshootingFigure 6-1 Digital Control Group Simplified Block Diagram

Página 51 - Resister) Setting

136 6 Digital Control TroubleshootingA1 CPU ReplacementA1 CPU ReplacementWhen you replace a faulty A1 CPU with a new one, remove the EEPROM from the

Página 52

6 137Digital Control TroubleshootingFIRMWARE INSTALLATION6. Digital ControlTroubleshootingFIRMWARE INSTALLATIONNo firmware is installed in new A1 CPU

Página 53

138 6 Digital Control TroubleshootingFIRMWARE INSTALLATIONFigure 6-4 Bootloader Menu Display4. Insert the firmware diskette into the floppy disk dri

Página 54

6 139Digital Control TroubleshootingSTART HERE6. Digital ControlTroubleshootingSTART HERE1. Check the Power-On SequenceSee the INSPECT THE POWER-ON S

Página 55

140 6 Digital Control TroubleshootingSTART HEREFigure 6-5 A1 Eight LEDs' Pattern

Página 56

6 141Digital Control TroubleshootingSTART HERE6. Digital ControlTroubleshooting2. Check the A1 CPU1. Turn the analyzer power on.2. Press [System], DI

Página 57 - Standard Resistor) Setting

Contents 71. General InformationOrganization of Service Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 58 - [Cal] - FIXTURE COMPEN - OPEN

142 6 Digital Control TroubleshootingSTART HEREFigure 6-6 Bootloader Display4. Check the A1 Volatile Memory1. Turn the analyzer power on.2. Press [S

Página 59

6 143Digital Control TroubleshootingSTART HERE6. Digital ControlTroubleshooting2. Press all of the front panel keys. The abbreviated key name should

Página 60 - Calculation Sheet

144 6 Digital Control TroubleshootingSTART HEREFigure 6-7 24 BIT I/O PORT Test Setup3. Turn the analyzer power on.4. Press [Preset], [System], DIAGN

Página 61

6 145Digital Control TroubleshootingSTART HERE6. Digital ControlTroubleshooting9. Press done.10. Connect a 10Base-T twisted pair “straight-through” c

Página 62

146 6 Digital Control TroubleshootingTROUBLESHOOT THE A51 GSP and A52 LCDTROUBLESHOOT THE A51 GSP and A52 LCDUse this procedure when the LCD (Liquid

Página 63

1477. Service Related Menus7 Service Related MenusThis chapter describes the functions of the diagnosis key menus. These menus are used to test, ver

Página 64

148 7 Service Related MenusDIAGNOSTIC TEST MenuDIAGNOSTIC TEST MenuThe diagnosis key menus consist of several menus that are accessed through the dia

Página 65

7 149Service Related MenusDIAGNOSTIC TEST Menu7. Service Related MenusFigure 7-1 Test Status on the DisplayTo see the test status of the desired tes

Página 66

150 7 Service Related MenusDIAGNOSTIC TEST MenuINTERNAL TESTThis group of tests run without external connections or operator interaction. All return

Página 67

7 151Service Related MenusDIAGNOSTIC TEST Menu7. Service Related Menus5: ADJUST DATA Verifies that all correction constants which compensate for the

Página 68

8 ContentsRequired Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 69

152 7 Service Related MenusDIAGNOSTIC TEST Menu(230kHz) signal at a frequency bus node. It checks that each measured value is within limits.15: A2/A9

Página 70

7 153Service Related MenusDIAGNOSTIC TEST Menu7. Service Related Menusthe adapter selection of the analyzer is changed to the “NONE” status.23: A2 AT

Página 71

154 7 Service Related MenusDIAGNOSTIC TEST Menu

Página 72

1558. Theory of Operation8 Theory of OperationThis chapter describes the general overall operation of the 4294A and the operation of each functional

Página 73

156 8 Theory of OperationOVERALL OPERATIONOVERALL OPERATIONOVERALL MEASUREMENT THEORYThe 4294A measures the impedance of the Device Under Test (DUT)

Página 74

8 157Theory of OperationOVERALL OPERATION8. Theory of OperationProcessor), the A52 LCD, and the A53 FDD. These assemblies combine to provide digital

Página 75

158 8 Theory of OperationPOWER SUPPLY OPERATIONPOWER SUPPLY OPERATIONThe power supply functional group consists of the following assemblies:• A40 Pre

Página 76 - Performance Test Record

8 159Theory of OperationPOWER SUPPLY OPERATION8. Theory of OperationFigure 8-2 Power Supply Functional Group, Simplified Block DiagramLine Power Mod

Página 77

160 8 Theory of OperationPOWER SUPPLY OPERATIONA40 PreregulatorThe A40 preregulator contains a rectifier and a switching regulator, converts the line

Página 78

8 161Theory of OperationDIGITAL CONTROL OPERATION8. Theory of OperationDIGITAL CONTROL OPERATIONThe digital control functional group consists of the

Página 79

Contents 95. Power Supply TroubleshootingINTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 80

162 8 Theory of OperationDIGITAL CONTROL OPERATIONFigure 8-3 Digital Control Group Simplified Block Diagram

Página 81

8 163Theory of OperationDIGITAL CONTROL OPERATION8. Theory of OperationA1 CPUThe A1 CPU consists of the following circuits and parts:CPU central proc

Página 82

164 8 Theory of OperationDIGITAL CONTROL OPERATIONterminal, the PRINTER port, the 24-BIT I/O PORT, and the LAN port.A51 GSPThe A51 GSP (graphics syst

Página 83

8 165Theory of OperationSOURCE THEORY8. Theory of OperationSOURCE THEORYThe source group generates all analog signals in the 4294A, a phase-locked te

Página 84

166 8 Theory of OperationSOURCE THEORY• 8MHz (for A/D converter clock): is supplied to the A8 through the A20.• 10MHz: is routed to the INT REF Outpu

Página 85

8 167Theory of OperationTRANSDUCER THEORY8. Theory of OperationTRANSDUCER THEORYThe transducer group balances the range resistor current with the DUT

Página 86

168 8 Theory of OperationTRANSDUCER THEORYA3 Hp Amp.The A3 Hp amp. senses the voltage across the DUT (Edut), and transmit the voltage (Edut) to the v

Página 87

8 169Theory of OperationVECTOR RATIO DETECTOR THEORY8. Theory of OperationVECTOR RATIO DETECTOR THEORYThe A8 board is the vector ratio detector. The

Página 88

170 8 Theory of OperationVECTOR RATIO DETECTOR THEORY

Página 90 - Warm-up for Adjustment

10 ContentsTest Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 91 - Order of Adjustment

1719. Parts Replacement9 Parts ReplacementThis chapter contains the 4294A’s replaceable parts list and the procedure to replace its assemblies.

Página 92 - Equipment Setup

172 9 Parts ReplacementReplaceable Part ListReplaceable Part ListOrdering InformationTo order part listed in the replaceable part lists, quote the Ag

Página 93

9 173Parts ReplacementReplaceable Part List9. Parts ReplacementParts ListFigure 9-1 Power Cables and Plug Configurations

Página 94

174 9 Parts ReplacementReplaceable Part ListFigure 9-2 Top View (Major Assemblies)

Página 95

Table 9-1 Top View (Major Assemblies)Reference DesignationAgilent Part NumberCheck DigitQty. DescriptionA2 04294-66502 3 1 High Current04294-69502 9

Página 96

176 9 Parts ReplacementReplaceable Part ListFigure 9-3 Top View (Cables and Wires)

Página 97

Table 9-2 Top View (Cables and Wires)Reference DesignationAgilent Part NumberCheck DigitQty. DescriptionA 04294-61601 3 1 RF Cable “A”, A3 to A4B 043

Página 98

178 9 Parts ReplacementReplaceable Part ListFigure 9-4 Top View (Miscellaneous Parts)Table 9-3 Top View (Miscellaneous Parts)Reference DesignationA

Página 99

9 179Parts ReplacementReplaceable Part List9. Parts ReplacementFigure 9-5 Top View (Shield)Table 9-4 Top View (Shield)Reference DesignationAgilent

Página 100 - Procedure

180 9 Parts ReplacementReplaceable Part ListFigure 9-6 Bottom View (Major Assemblies)Table 9-5 Bottom View (Major Assembly)Reference DesignationAgi

Página 101

Contents 11A4 Lc Amplifier Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 102

9 181Parts ReplacementReplaceable Part List9. Parts ReplacementFigure 9-7 Bottom View (Cables and Wires)Table 9-6 Bottom View (Cables and Wires)Ref

Página 103

182 9 Parts ReplacementReplaceable Part ListFigure 9-8 Bottom View (Miscellaneous Parts)

Página 104

Table 9-7 Bottom View (Miscellaneous Parts)Reference DesignationAgilent Part NumberCheck DigitQty. Description1 0515-1719 3 4 Screw M42 0515-1550 0 1

Página 105

184 9 Parts ReplacementReplaceable Part ListFigure 9-9 Front Assembly 1Table 9-8 Front Assembly 1Reference DesignationAgilent Part NumberCheck Digi

Página 106

9 185Parts ReplacementReplaceable Part List9. Parts ReplacementFigure 9-10 Front Assembly 2Table 9-9 Front Assembly 2Reference DesignationAgilent P

Página 107

186 9 Parts ReplacementReplaceable Part ListFigure 9-11 Front Assembly 3Table 9-10 Front Assembly 3Reference DesignationAgilent Part NumberCheck Di

Página 108

9 187Parts ReplacementReplaceable Part List9. Parts ReplacementFigure 9-12 Front Assembly 4Tabl e 9 - 11 Front Ass emb l y 4Reference Designation

Página 109

188 9 Parts ReplacementReplaceable Part ListFigure 9-13 Front Assembly 5Table 9-12 Front Assembly 5Reference DesignationAgilent Part NumberCheck Di

Página 110 - Adjustment

9 189Parts ReplacementReplaceable Part List9. Parts ReplacementFigure 9-14 Front Assembly 6Table 9-13 Front Assembly 6Reference DesignationAgilent

Página 111

190 9 Parts ReplacementReplaceable Part ListFigure 9-15 Front Assembly 7Table 9-14 Front Assembly 7Reference DesignationAgilent Part NumberCheck Di

Página 112

○C Copyright 2009 Agilent Technologies Change 2 Change the Measurement Uncertainty value of OSC Level Accuracy Test (Page 69 to 70) to the

Página 114 - TROUBLESHOOTING SUMMARY

9 191Parts ReplacementReplaceable Part List9. Parts ReplacementFigure 9-16 Front Assembly 8Table 9-15 Front Assembly 8Reference DesignationAgilent

Página 115 - Troubleshooting

192 9 Parts ReplacementReplaceable Part ListFigure 9-17 Front Assembly 9Table 9-16 Front Assembly 9Reference DesignationAgilent Part NumberCheck Di

Página 116 - START HERE

9 193Parts ReplacementReplaceable Part List9. Parts ReplacementFigure 9-18 Front Assembly 10Table 9-17 Front Assembly 10Reference DesignationAgilen

Página 117 - INSPECT THE POWER ON SEQUENCE

194 9 Parts ReplacementReplaceable Part ListFigure 9-19 Rear Assembly 1Table 9-18 Rear Assembly 1Reference DesignationAgilent Part NumberCheck Digi

Página 118 - Check the Parallel Interface

9 195Parts ReplacementReplaceable Part List9. Parts ReplacementFigure 9-20 Rear Assembly 2Table 9-19 Rear Assembly 2Reference DesignationAgilent Pa

Página 119 - Execute the Individual Test

196 9 Parts ReplacementReplaceable Part ListFigure 9-21 Rear Assembly 3

Página 120

Table 9-20 Rear Assembly 3Reference DesignationAgilent Part NumberCheck DigitQty. Description1 1250-0252 6 1 BNC Connector2190-0102 8 1 Washer2950-00

Página 121

198 9 Parts ReplacementReplaceable Part ListFigure 9-22 Rear Assembly 4Table 9-21 Rear Assembly 4Reference DesignationAgilent Part NumberCheck Digi

Página 122

9 199Parts ReplacementReplaceable Part List9. Parts ReplacementFigure 9-23 Rear Assembly 5Table 9-22 Rear Assembly 5Reference DesignationAgilent Pa

Página 123 - Null Detector Offset CC

200 9 Parts ReplacementReplaceable Part ListFigure 9-24 Side ViewTable 9-23 Side ViewReference DesignationAgilent Part NumberCheck DigitQty. Descri

Página 124

131. General Information1 General InformationThis Service Manual is a guide to servicing the 4294A Precision Impedance Analyzer. The Service Manual

Página 125 - 5. Power Supply

9 201Parts ReplacementReplaceable Part List9. Parts ReplacementFigure 9-25 Chassis AssemblyTable 9-24 Chassis AssemblyReference DesignationAgilent

Página 126 - INTRODUCTION

Table 9-25 Other PartsReference DesignationAgilent Part NumberCheck DigitQty. Description1 04294-61001 7 1 100 Ω Resistor2 1250-1859 1 1 BNC(m)-BNC(m

Página 127

9 203Parts ReplacementReplacement Procedure9. Parts ReplacementReplacement ProcedureThis Chapter describes how to replace the 4294A major assemblies.

Página 128

204 9 Parts ReplacementReplacement Procedure2. Remove the two screws at the side strap handle caps to remove the strap.3. Slide off the side cover to

Página 129 - 4. Check the A1 +5 VD LED

9 205Parts ReplacementReplacement Procedure9. Parts Replacement6. Gradually pull the rear panel assembly out from the rear frame.A1 CPU Board Replace

Página 130 - 5. Check the A9 Two LEDs

206 9 Parts ReplacementReplacement Procedure6. Disconnect the cable “C” from the A3 Hp Amplifier, and connect it to the replacement A3.A4 Lc Amplifie

Página 131

9 207Parts ReplacementReplacement Procedure9. Parts Replacement2. Remove the top shield.3. Disconnect the cables “K” and “L” from the A7 board.4. Dis

Página 132 - Power Supply Troubleshooting

208 9 Parts ReplacementReplacement ProcedureFigure 9-26 A9 Floating Power Supply3. Disconnect the cables (item ¦ and Æ)from the A9 board.4. Remove f

Página 133

9 209Parts ReplacementReplacement Procedure9. Parts ReplacementA52 LCD RemovalTools Required• Torx screwdriver, T15• Pozidriv Screwdriver, pt size #1

Página 134 - 3. Check the A50 SHUTDOWN LED

210 9 Parts ReplacementReplacement ProcedureFigure 9-27 Preregulator ReplacementPower Switch ReplacementTools Required• Torx screwdriver, T15• Pozid

Página 135 - 3. Remove Assemblies

14 1 General InformationOrganization of Service ManualOrganization of Service ManualThis manual consists of the major chapters listed below. This sec

Página 136 - STEADILY

9 211Parts ReplacementReplacement Procedure9. Parts ReplacementFigure 9-28 Cable Clamp3. Remove the angel from the chassis.4. Disconnect the power s

Página 137

212 9 Parts ReplacementReplacement Procedure

Página 138

21310. Post Repair Procedures10 Post Repair ProceduresThis chapter lists the procedures required to verify the 4294A operation after an assembly is

Página 139 - CONVERTER

214 10 Post Repair ProceduresPOST REPAIR PROCEDURESPOST REPAIR PROCEDURESTable 10-1 Post Repair Procedures lists the required procedures that must be

Página 140 - Table 5-1 A50 Power Supplies

10 215Post Repair ProceduresPOST REPAIR PROCEDURES10. Post Repair ProceduresA7 Synthesizer Frequency Accuracy CC Image Rejection CC Null Detector Off

Página 141 - 6. Digital Control

216 10 Post Repair ProceduresPOST REPAIR PROCEDURES

Página 142

217A. Manual ChangesA Manual ChangesThis appendix contains the information required to adapt this manual to earlier versions or configurations of th

Página 143

218 Appendix A Manual ChangesManual ChangesManual ChangesTo adapt this manual to your 4294A, refer to Table A-1 and Table A-2.Tabl e A -1

Página 144 - A1 CPU Replacement

219B. Power RequirementB Power Requirement

Página 145 - FIRMWARE INSTALLATION

220 Appendix B Power RequirementReplacing FuseReplacing FuseFuse SelectionSelect proper fuse according to the Table B-1.Table B-1 Fuse SelectionFuse

Página 146

1 15General InformationInstruments Covered by This Manual1. General InformationInstruments Covered by This ManualAgilent Technologies uses a two-part

Página 147

Appendix B 221Power RequirementPower RequirementsB. Power RequirementPower RequirementsThe 4294A requires the following power source:Voltage : 90 to

Página 148

222 Appendix B Power RequirementPower RequirementsFigure B-1 Power Cable Supplied

Página 149 - 2. Check the A1 CPU

223C. Error MessagesC Error messagesThe 4294A provides error messages to indicate its operating status. This appendix describes the error messages o

Página 150

224 Appendix C Error messagesOrder of AlphabetNOTE If an error with a positive error number occurs, only its error message is displayed on the LCD fo

Página 151 - 8. Check the 24 BIT I/O PORT

Appendix C 225Error messagesOrder of AlphabetC. Error Messages136 BRIDGE UNBALANCEDBecause the connection of a DUT or the setting for it was incorrec

Página 152 - 9. Check the LAN Operation

226 Appendix C Error messagesOrder of Alphabeterror coefficients based on the measurement.-281 Cannot create programPrograms cannot be created. Insuf

Página 153

Appendix C 227Error messagesOrder of AlphabetC. Error MessagesA comprehensive syntax error occurred for which the 4294A could not detect further deta

Página 154

228 Appendix C Error messagesOrder of Alphabet138 DCBIAS CONSTANT OPERATION FAILEDWhen the DC bias mode was constant voltage (VOLT CONSTANT) or const

Página 155 - 7 Service Related Menus

Appendix C 229Error messagesOrder of AlphabetC. Error MessagesAn equation data element was received at where the 4294A did not accept any equation da

Página 156 - DIAGNOSTIC TEST Menu

230 Appendix C Error messagesOrder of Alphabet-105 GET not allowedA group execution trigger (GET) was received in a program message. (Refer to IEEE48

Página 157 - Test Descriptions

16 1 General InformationRequire EquipmentRequire EquipmentTable 1-1 lists the recommended equipment for performing maintenance on the 4294A.Table 1-

Página 158

Appendix C 231Error messagesOrder of AlphabetC. Error Messages-161 Invalid block dataBlock data was expected but given block data was invalid for som

Página 159

232 Appendix C Error messagesOrder of AlphabetL132 LIST SWEEP ONLYWhen a sweep type other than the list sweep was specified, you attempted to execute

Página 160

Appendix C 233Error messagesOrder of AlphabetC. Error MessagesWhen the memory trace was not displayed, you attempted to set the function that needed

Página 161

234 Appendix C Error messagesOrder of Alphabet(COPY FILE) key, no file or directory corresponding to the execution of the key was on the floppy disk.

Página 162 - Service Related Menus

Appendix C 235Error messagesOrder of AlphabetC. Error MessagesWhen the percent format was not selected as the display format, you attempted to select

Página 163 - 8 Theory of Operation

236 Appendix C Error messagesOrder of Alphabetthe error. This code simply indicates the occurrence of a Query error that is defined in IEEE488.2,11.5

Página 164 - OVERALL OPERATION

Appendix C 237Error messagesOrder of AlphabetC. Error MessagesWhen the band width function was off, you attempted to execute a command to search the

Página 165

238 Appendix C Error messagesOrder of AlphabetBefore executing a command to set the phase unit or a command to set the phase expansion display, perfo

Página 166 - POWER SUPPLY OPERATION

Appendix C 239Error messagesOrder of AlphabetC. Error MessagesFor a 24-bit input/output (I/O) shared port (port C and port D), you attempted to execu

Página 167 - Line Power Module

240 Appendix C Error messagesOrder of Alphabet

Página 168 - A9 Floating Power Supply

1 17General InformationRequire Equipment1. General InformationAdapter 7 mm-N(m) AdapterDual Banana-BNC(f) AdapterN(m)-BNC(m) AdapterBNC(f)-BNC(f) Ada

Página 169 - DIGITAL CONTROL OPERATION

18 1 General InformationRequire Equipment

Página 170 - Theory of Operation

192. Performance Test2 Performance TestThis chapter contains the performance test for the 4294A Precision Impedance Analyzer. These performance test

Página 171 - A34 I/O Connector

20 2 Performance TestPerformance TestPerformance TestIntroductionThis section provides the test procedures used to verify that the 4294A’s specificat

Página 172 - A54 Inverter

2 21Performance TestPerformance Test2. Performance TestFrequency Accuracy TestThis test checks the frequency accuracy of the internal frequency refer

Página 173 - SOURCE THEORY

○C Copyright 2009 Agilent Technologies OSC Level Accuracy (with 42942A Terminal Adapter) OSC Level Frequency Test Limits Test Result Mea

Página 174 - A2 Hc Amp

22 2 Performance TestPerformance TestFigure 2-1 Frequency Accuracy Test Setup2. Initialize the frequency counter. Then set the controls as follows.G

Página 175 - TRANSDUCER THEORY

2 23Performance TestPerformance Test2. Performance TestOSC Level Accuracy TestThis test checks the actual power level of the test signal at 10 MHz.Sp

Página 176 - A6 Vector Generator

24 2 Performance TestPerformance TestFigure 2-2 OSC Level Accuracy Test Setup 12. Press Reset key (blue,⇒) to initialize the multimeter, then Set it

Página 177 - VECTOR RATIO DETECTOR THEORY

2 25Performance TestPerformance Test2. Performance Test7. Confirm that the monitor accuracy is within the monitor accuracy limit.NOTE If the monitor

Página 178

26 2 Performance TestPerformance TestFigure 2-3 OSC Level Accuracy Test Setup 212. Set the 4294A as follows.Setting OperationCenter Frequency: 1 MHz

Página 179

2 27Performance TestPerformance Test2. Performance Test16. Repeat from Step 13 to 15 for all frequency settings in Table 2-2Table 2-2 OSC Level Accu

Página 180 - 9 Parts Replacement

28 2 Performance TestPerformance TestFigure 2-4 42942A Setup25. Setup the 42942A as follows.a. Press [Cal],Adapter,7mm 42942A,SETUP.b. Connect the O

Página 181 - Replaceable Part List

2 29Performance TestPerformance Test2. Performance TestFigure 2-5 OSC Level Accuracy Test Setup 327. Set the 4294A as follows.Setting OperationCente

Página 182 - Parts List

30 2 Performance TestPerformance Test31. Repeat from Step 28 to 30 for all frequency settings in Table 2-3Table 2-3 OSC Level Accuracy Test (Interfa

Página 183 - Parts Replacement

2 31Performance TestPerformance Test2. Performance TestDC Bias Monitor Accuracy TestThis test checks the accuracy of the DC bias monitor.Specificatio

Página 184

○C Copyright 2009 Agilent Technologies Accessory: Nothing Standard: 1pF OSC Level Frequency Parameter Test Limits Test Result Measurement U

Página 185

32 2 Performance TestPerformance Testb. Press [NPLC] - [1] - [0] - [0] - [Enter].3. Press [Preset] to initialize the 4294A, then set it as follows.Se

Página 186

2 33Performance TestPerformance Test2. Performance TestFigure 2-7 DC Bias Monitor Accuracy Test (current) Setup 9. Press Reset key (blue - ⇒) to ini

Página 187

34 2 Performance TestPerformance Test12. Press [Trigger] - Single for a single sweep measurement.13. Record the multimeter reading and the DC Bias mo

Página 188 - Table 9-4 Top View (Shield)

2 35Performance TestPerformance Test2. Performance TestDC Bias Level Accuracy TestThis test checks the accuracy of the DC bias level at several level

Página 189 - 4TP Assembly

36 2 Performance TestPerformance TestFigure 2-8 DC Bias Level Accuracy Test (voltage) Setup2. Press Reset key (blue - ⇒) to initialize the multimete

Página 190

Table 2-6 DC Bias Monitor Accuracy Test (voltage) SettingBias Level0 V25 V40 V–25 V–40 V2 37Performance TestPerformance Test2. Performance Test8. Di

Página 191

38 2 Performance TestPerformance TestMeasurement Accuracy TestThis test checks the DC bias level of the stimulus signal at several frequencies from 4

Página 192

2 39Performance TestPerformance Test2. Performance TestFigure 2-9 Measurement Accuracy Test Setup (Accessory: None)5. Press [Cal] - FIXTURE COMPEN -

Página 193 - Table 9-8 Front Assembly 1

40 2 Performance TestPerformance TestFigure 2-10 Sweep List for Measurement Accuracy Testc. Press done.10. Press done to save the sweep list menu.11

Página 194 - Table 9-9 Front Assembly 2

2 41Performance TestPerformance Test2. Performance TestFigure 2-11 Measurement Value List15. Record the 4294A reading on the calculation sheet for a

Página 195 - Table 9-10 Front Assembly 3

○C Copyright 2009 Agilent Technologies Accessory: Nothing Standard: 1000 pF OSC Level Frequency Parameter Test LimitsTest Result Measuremen

Página 196

Table 2-8 Measurement Accuracy Test(Accessory: None, Standard Capacitor) SettingStandard CapacitorOSC Level Frequency1 pF 0.5 V 1 MHz3 MHz10 pF 0.5 V

Página 197 - Table 9-12 Front Assembly 5

2 43Performance TestPerformance Test2. Performance Test16. Connect the 10 Ω standard resistor to the 4294A UNKNOWN Terminal as shown in Figure 2-9 on

Página 198 - Table 9-13 Front Assembly 6

44 2 Performance TestPerformance Test21. Press [Cal] - Adapter - 7mm 42942A to set the adapter to the 42942A.22. Setup the 4294A as follows.NOTE If S

Página 199 - Table 9-14 Front Assembly 7

2 45Performance TestPerformance Test2. Performance Testc. Gently inserts the airline center conductor into the open termination center conductor. Mat

Página 200 - Table 9-15 Front Assembly 8

46 2 Performance TestPerformance Testtorque. (A 1/2 inch open end wrench may be necessary to hold the airline stationary.)Figure 2-14 10 cm Airline

Página 201 - Table 9-16 Front Assembly 9

2 47Performance TestPerformance Test2. Performance TestFigure 2-15 Connection for phase compensationd. Press PHASE COMP to measure the phase compens

Página 202 - Table 9-17 Front Assembly 10

48 2 Performance TestPerformance TestFigure 2-17 Measurement Accuracy Test Setup (Accessory: 1m Test Leads)35. Connect the OPEN Termination as shown

Página 203 - Table 9-18 Rear Assembly 1

Table 2-12 Measurement Accuracy Test(Accesory:1m Test Leads, Standard Capacitor) SettingStandard CapacitorOSC Level Frequency10 pF 0.5 V 1 kHz1 MHz3

Página 204 - Table 9-19 Rear Assembly 2

50 2 Performance TestPerformance TestFigure 2-18 Measurement Accuracy Test Setup (Accessory: 2m Test Leads)49. Press [Cal] - FIXTURE COMPEN - OPEN50

Página 205 - Figure 9-21 Rear Assembly 3

Table 2-14 Measurement Accuracy Test(Accesory:2m Test Leads, Standard Capacitor) SettingStandard CapacitorOSC Level Frequency10 pF 0.5 V 1 kHz1 MHz3

Página 206 - Table 9-20 Rear Assembly 3

○C Copyright 2009 Agilent Technologies Accessory: Nothing Standard: 10 Ω OSC Level Frequency Parameter Test Limits Test Result Measurement

Página 207 - Table 9-21 Rear Assembly 4

52 2 Performance TestCalculation SheetCalculation SheetOSC Level Accuracy TestOSC Level Accuracy(Accessory: Nothing, 40 Hz to 100 kHz)OSCLevel [a]Fre

Página 208 - Table 9-22 Rear Assembly 5

2 53Performance TestCalculation Sheet2. Performance TestOSC Level Accuracy (Accessory: Nothing, 1 MHz to 110 MHz)OSCLevel [a]Freq. MonitorReading[b]P

Página 209 - Table 9-23 Side View

54 2 Performance TestCalculation SheetDC Bias Monitor Accuracy TestDC Bias Monitor Accuracy Test (voltage)Bias Level Monitor Reading [a]Multimeter Re

Página 210 - Table 9-24 Chassis Assembly

2 55Performance TestCalculation Sheet2. Performance TestDC Bias Level Accuracy TestDC Bias Level Accuracy Test (voltage)Bias Level [a] Multimeter Rea

Página 211 - Table 9-25 Other Parts

56 2 Performance TestCalculation SheetMeasurement Accuracy TestStandard Calibration Values (16380A/C,42030A)Standard Frequency Parameter Calibration

Página 212 - Replacement Procedure

2 57Performance TestCalculation Sheet2. Performance Test0.01 μF 120 Hz Cp nF CV23D

Página 213 - Rear Panel Removal

58 2 Performance TestCalculation SheetStandard Calibration Values (16190A)Table 2-16 Standard Calibration ValueStandard Frequency Parameter Calibrat

Página 214 - A3 Hp Amplifier Replacement

2 59Performance TestCalculation Sheet2. Performance TestMeasurement Accuracy TestAccessory: NothingStandard: 1 pFOSC Level Frequency Parameter 42

Página 215 - A4 Lc Amplifier Replacement

60 2 Performance TestCalculation SheetAccessory: NothingStandard: 100 pFOSC Level Frequency Parameter 4294A reading [a]Test Result Equation0.5 V

Página 216 - A7 Synthesizer Replacement

2 61Performance TestCalculation Sheet2. Performance TestAccessory: NothingStandard: 1000 pFOSC Level Frequency Parameter 4294A reading [a]Test Re

Página 217 - A53 FDD Removal

○C Copyright 2009 Agilent Technologies Accessory: 42942A Terminal Adapter Standard: 10 cm Airline with SHORT OSC Level Frequency Parameter

Página 218 - A40 Preregulator Replacement

62 2 Performance TestCalculation SheetAccessory: NothingStandard: 0.1 μFOSC Level Frequency Parameter 4294A reading [a]Test Result Equation0.5 V

Página 219 - Power Switch Replacement

2 63Performance TestCalculation Sheet2. Performance TestAccessory: NothingStandard: 1 ΩOSC Level Frequency Parameter 4294A reading [a]Test Result

Página 220

64 2 Performance TestCalculation SheetAccessory: 42942A Terminal AdapterStandard: OPENOSC Level Frequency Parameter 4294A reading [a]Test Result

Página 221

2 65Performance TestCalculation Sheet2. Performance TestAccessory: 1m Test LeadsStandard: 100 pFOSC Level Frequency Parameter 4294A reading [a]Te

Página 222 - 10 Post Repair Procedures

66 2 Performance TestCalculation SheetAccessory: 2m Test LeadsStandard: 10 pFOSC Level Frequency Parameter 4294A reading [a]Test Result Equation0

Página 223 - POST REPAIR PROCEDURES

2 67Performance TestCalculation Sheet2. Performance TestAccessory: 2m Test LeadsStandard: 100 mΩOSC Level Frequency Parameter 4294A reading [a]Te

Página 224

68 2 Performance TestPerformance Test RecordPerformance Test RecordAgilent Technologies 4294A Precision Impedance AnalyzerSerial Number:

Página 225 - Post Repair Procedures

2 69Performance TestPerformance Test Record2. Performance TestOSC Level Accuracy TestOSC Level Accuracy (Accessory: Nothing)OSC Level Frequency Test

Página 226 - A Manual Changes

70 2 Performance TestPerformance Test RecordOSC Level Accuracy (with 42942A Terminal Adapter)OSC Level Frequency Test Li mi t Test Result Measurement

Página 227 - Manual Changes

2 71Performance TestPerformance Test Record2. Performance TestDC Bias Level Accuracy TestBias Level Test Limit Test R e sul t Measurement Uncertainty

Página 228 - B Power Requirement

○C Copyright 2009 Agilent Technologies Accessory: 2m Test Leads Standard: 10 pF OSC Level Frequency Parameter Test Limits Test Result Meas

Página 229 - Replacing Fuse

72 2 Performance TestPerformance Test RecordMeasurement Accuracy TestAccessory: NothingStandard: 1 pFOSC Level Frequency Parameter Test Limit Te

Página 230 - Power Requirements

2 73Performance TestPerformance Test Record2. Performance TestAccessory: NothingStandard: 100 pFOSC Level Frequency Parameter Test Limit Te st R

Página 231 - Power Requirement

74 2 Performance TestPerformance Test RecordAccessory: NothingStandard: 1000 pFOSC Level Frequency Parameter Test Limit Te st Re sult Measuremen

Página 232 - C Error messages

2 75Performance TestPerformance Test Record2. Performance TestAccessory: NothingStandard: 0.1 μFOSC Level Frequency Parameter Test Limit Te st R

Página 233 - Order of Alphabet

76 2 Performance TestPerformance Test RecordAccessory: NothingStandard: 10 ΩOSC Level Frequency Parameter Test Limit Te st Re sult Measurement U

Página 234

2 77Performance TestPerformance Test Record2. Performance TestAccessory: 42942A Terminal AdapterStandard: 50 ΩOSC Level Frequency Parameter Test

Página 235

78 2 Performance TestPerformance Test RecordAccessory: 1m Test LeadsStandard: 10 pFOSC Level Frequency Parameter Test Limit Te st Re sult Measur

Página 236

2 79Performance TestPerformance Test Record2. Performance TestAccessory: 1m Test LeadsStandard: 1 ΩOSC Level Frequency Parameter Test Limit Te s

Página 237

80 2 Performance TestPerformance Test RecordAccessory: 2m Test LeadsStandard: 100 pFOSC Level Frequency Parameter Test Limit Te st Re sult Measu

Página 238

811. Chapter Title2. Chapter Title 3. Adjustment4. Chapter Title5. Chapter Title3 AdjustmentThis Chapter describes the Adjustment procedures require

Página 239

Agilent 4294A Precision Impedance AnalyzerService ManualThird EditionManufacturing No. 04294-90110January 2007

Página 240

82 3 AdjustmentSafety ConsiderationsSafety ConsiderationsThis manual contains NOTEs, CAUTIONs, and WARNINGs that must be followed to ensure the safet

Página 241

3 83AdjustmentOrder of Adjustment1. Chapter Title2. Chapter Title 3. Adjustment4. Chapter Title5. Chapter TitleOrder of AdjustmentWhen performing mor

Página 242

84 3 AdjustmentPreparation for using the Adjustment ProgramPreparation for using the Adjustment ProgramTo use the Adjustment Program, some preparatio

Página 243

3 85AdjustmentPreparation for using the Adjustment Program1. Chapter Title2. Chapter Title 3. Adjustment4. Chapter Title5. Chapter TitleFigure 3-1 A

Página 244

86 3 AdjustmentRunning the Adjustment ProgramRunning the Adjustment Program1. Start the HP VEE.2. Load the adjustment program file into the HP VEE as

Página 245

3 87AdjustmentFrequency Accuracy Correction Constants1. Chapter Title2. Chapter Title 3. Adjustment4. Chapter Title5. Chapter TitleFrequency Accuracy

Página 246

88 3 AdjustmentImage Rejection Correction ConstantsImage Rejection Correction ConstantsThe purpose of this procedure is to generate the correction co

Página 247

3 89AdjustmentLPOT-Mixer Offset Correction Constants1. Chapter Title2. Chapter Title 3. Adjustment4. Chapter Title5. Chapter TitleLPOT-Mixer Offset C

Página 248

90 3 AdjustmentNull Detector Offset Correction ConstantsNull Detector Offset Correction ConstantsThe purpose of this procedure is to adjust the Null

Página 249 - Error messages

3 91AdjustmentNull Loop Phase Correction Constants1. Chapter Title2. Chapter Title 3. Adjustment4. Chapter Title5. Chapter TitleNull Loop Phase Corre