Agilent Technologies 8935 Series Especificações Página 329

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 328
328 Appendix A
Error Messages
Flash ROM Firmware Upgrade Error Messages
Programming Voltage Error
The programming voltage is supplied to the flash ROMs from the power
supply through the Controller assembly. The fault is most likely on the
Controller but can be caused by the Filter/Regulator assembly.
ROM Checksum Error
With the new firmware loaded into the Test Set’s flash ROMs, the
checksum on the ROM is tested. A faulty checksum is most likely
caused by the flash ROMs themselves or possibly the controller circuits.
In either case replace the A2A31 Controller assembly.
Vista de página 328
1 2 ... 324 325 326 327 328 329 330 331 332 333 334 ... 342 343

Comentários a estes Manuais

Sem comentários