
228 Chapter 8
Performance Tests
CDMA Analyzer Channel Power Level Accuracy Performance Test 28
2. On the signal generator:
a. Set the frequency to 881.32 MHz.
b. Set the output level so the measuring receiver reads 11 dBm
3. Push
Preset on the Test Set.
4. On the Test Set:
a. Change the Avg Pwr field to Chan Pwr.
b. Set the Tune Freq to 881.32 MHz
c. Select Calibrate under Chn Pwr Cal.
5. Record the Chan Pwr measurement in the PTR, see Table 9-29,
“CDMA Analyzer Channel Power Level Accuracy Test 28 Record,” on
page 280.
6. Repeat steps 2 and 3 for each of the data points listed in the PTR.
NOTE
The Chan Pwr Cal is only required when the frequency is changed.
Comentários a estes Manuais