Agilent Technologies 8935 Series Especificações Página 218

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 217
Chapter 8 217
Performance Tests
RF Analyzer Residual FM Performance Test 23
RF Analyzer Residual FM
Performance Test 23
The purpose of this test is to verify that the Test Set meets the
specification limits in PTR (Performance Test Record) Table 9-23, “RF
Analyzer Residual FM Test 23 Record,” on page 274. The AM/FM test
source provides a CW signal with minimal residual FM. The FM is
measured by the Test Set’s internal RF analyzer.
Figure 8-28 Setup
Procedure
1. On the AM/FM test source, set the test mode to residual FM.
2. On the Test Set:
a. Press
Preset.
b. Select the INSTRUMENT CONFIGURE screen.
c. Set the RF Display field to Freq.
d. Select the CDMA GENERATOR screen.
e. Set the CW RF Path field to Bypass.
f. Select the RF ANALYZER screen.
g. Set the Tune Freq to 560 MHz.
h. Set the Input Port field to Ant.
i. Set the IF Filter field to 230 kHz.
j. Select the AF ANALYZER screen.
k. Set the AF Anl In field to FM Demod.
l. Set the Filter 1 field to 300Hz HPF.
m.Set the Filter 2 field to 3kHz LPF.
n. Set the Detector field to RMS.
ANT IN
AM/FM Test Source
LOW RESIDUAL OUTPUT
Vista de página 217