212 Chapter 8
Performance Tests
RF Analyzer FM Accuracy Performance Test 20
3. On the audio analyzer:
a. Reset the instrument.
b. Set the output frequency to 50 Hz.
4. On the Test Set:
a. Press
Preset.
b. Select the INSTRUMENT CONFIGURE screen.
c. Set the RF Display field to Freq.
d. Select the CDMA GENERATOR screen.
e. Set the CW RF Path field to Bypass.
f. Select the RF ANALYZER screen.
g. Set the Tune Freq to 12.5 MHz.
h. Set the Input Port field to Ant.
i. Set the IF Filter field to 230 kHz.
j. Select the AF ANALYZER screen.
k. Set the AF Anl In field to FM Demod.
l. Set the Filter 2 field to >99kHz LP.
m.Set the Detector field to RMS.
5. For each RF output from the AM/FM test source (12.5 MHz and 400
MHz corresponding to the FM÷32 and FM outputs) shown in the
Performance Test Record (PTR), do the following:
a. Set the audio analyzer’s frequency (rate) as shown in the PTR.
b. Adjust the audio analyzer’s level until the measuring receiver
reads the FM deviation shown in the PTR.
c. Read the FM deviation on the Test Set and compare the results to
the limits shown in the PTR.