Chapter 8 195
Performance Tests
RF Generator Harmonics Spectral Purity Performance Test 6
RF Generator Harmonics Spectral Purity
Performance Test 6
The purpose of this test is to verify that the Test Set meets the
specification limits in PTR (Performance Test Record) Table 9-6, “RF
Generator Harmonics Spectral Purity Test 6 Record,” on page 250.
Harmonic signals with the carrier set to several frequencies and two
different levels (maximum output and minimum level vernier) are
searched for by an RF spectrum analyzer.
Figure 8-11 Setup
Procedure
1. Set up the spectrum analyzer in accordance with its operating
manual.
2. On the Test Set:
a. Press
Preset.
b. Select the INSTRUMENT CONFIGURE screen.
c. Set the RF Display field to Freq.
d. Select the CDMA GENERATOR screen.
e. Set the CW RF Path field to Bypass.
f. Select the RF GENERATOR screen.
g. Set the RF Gen Freq to 1 MHz.
h. Set the Amplitude to -10 dBm.
3. Set the test set’s RF generator to the frequencies and levels shown in
the Performance Test Record (PTR) and measure the second and
third harmonics. For each measurement convert the harmonic level
to dB below the fundamental (dBc) and compare the computed levels
to the limits.
INPUT 50Ω
DUPLEX OUT
Spectrum Analyzer