72 Chapter 3
Troubleshooting
AF, RF, & CDMA Diagnostics (Step 3)
CDMA Diagnostics
This program tests the local oscillators and the power supplies of the
following assemblies:
A2A120 LO IF/IQ Modulator - LO IF portion only
A2A100 CDMA Generator Reference
To run CDMA diagnostics, see “CDMA Diagnostics” on page 72.
NOTE
Before ordering a replacement assembly...
Before ordering an assembly based on the results of the diagnostics, you
should verify the diagnostics by other means if possible. This could
include using manual troubleshooting procedures and descriptions of
the AF, RF, and CDMA diagnostics in this chapter, and/or block
diagrams in Chapter 10 , “Block Diagrams,” on page 283. If you still
lack confidence in troubleshooting or diagnosing the problem or faulty
assembly, call “Factory Support” on page 46, for troubleshooting
assistance.
Figure 3-15 CDMA Diagnostics Screen
SERVICE MENU
1
2
3
4
5
Help
Exit
Functional Diagnostics
AF Diagnostics
RF Diagnostics
CDMA Diagnostics
Edit RF Diagnostic Limits
Periodic Calibration
IQ Calibration
Eb/No Calibration
=>
Move pointer to desired program using the
knob then press the knob. Press Help for
information on the tests. Press Exit to abort.
cdmascrn.eps
SERVICE MENU Screen
CDMA Diagnostics
Screen
INSTRUCTIONS
1
2
3
4
5
Resume
Serv Menu
Exit
Connect DUPLEX OUT to ANT IN with a short cable.
Select Resume to continue, Serv Menu to go to the
Service Menu, Exit to abort.