42 Chapter 2
Product Information
Upgrades
4. Follow the instructions on the screen.
5. After the firmware is downloaded, be sure to cycle the Test Set’s
power off-and-on to complete this procedure.
Load the DSP Firmware
The following procedure loads the digital signal processor firmware.
1. Press the
Preset key.
2. Insert the E6380-10002 memory card into the Test Set.
3. Press the Menu key.
4. Set the Select Procedure Location: field to Card.
5. Set the Select Procedure Filename field to DLFIRM.
6. Select Run Test (
k1).
7. Follow the instructions on the screen.
After the firmware is downloaded, be sure to cycle the Test System’s
power.
CDMA 2000 Enable Process
The following procedure will allow you to enable the CDMA 2000
feature for the 8935 CDMA Base Station Test Set, (E6380A option 200):
1. Verify the test set is turned off.
2. Insert the Host FW card 1 of 2 into the PCMCIA card slot on the
front of the test set.
3. Power the unit on. Follow the on screen directions to start the Host
FW download procedure.
4. Once the second card is loaded, select the Menu screen.
5. Insert the DSP card 1of 1 into the PCMCIA card slot.
6. On the Menu screen, choose Card for Procedure Location.
7. Select DLFIRM for Procedure Filename.
8. Load and run the program by pressing the Run Test soft key, K1.
9. Once the DSP code is loaded, remove the DSP card and cycle power.
10.Insert the Config. Program card into the PCMCIA slot.
11.From the Menu screen select Card location and Config Filename.
12.Load and run the program by pressing the Run Test softkey, K1.
13.Follow the on screen direction to enable Option 200 (IS2000_CDMA).