84 Chapter 3
Troubleshooting
Manual Troubleshooting Procedures (Step 4)
3. Set the RF signal generator’s frequency to 100, then 500, 900, and
1800 MHz. For each frequency reset the Tune Freq to that
frequency. The Test Sets measurements should read as follows:
a. TX Power should read approximately 0.001 W for each frequency.
b. Frequency should read 100, 500, 900, and 1800 MHz respectively.
c. Press the
Spec Anl key to access the analog spectrum analyzer.
Observe the level and frequency of the signal.
Isolating the RF Source
The RF Generator function uses the following assemblies. Refer to
Figure 3-19 and the block diagrams in Chapter 10 , “Block Diagrams,”
on page 283.
A2A120 LO IF/IQ Modulator
A2A25 Signal Generator Synthesizer
A2A24 Output Section
A2A110 Upconverter
Figure 3-19 Isolating the RF Source