Chapter 8 229
Performance Tests
CDMA Analyzer Modulation Accuracy Performance Test 29
CDMA Analyzer Modulation Accuracy
Performance Test 29
The purpose of this test is to verify that the Test Set meets the
specification limits in PTR (Performance Test Record) Table 9-29,
“CDMA Analyzer Channel Power Level Accuracy Test 28 Record,” on
page 280. This test verifies the performance of the Test Set’s CDMA
analyzer. The CDMA analyzer is performing properly if the rho values
calculated as the results of the following procedure fall within the lower
and upper limits specified in Table 9-29 on page 280.
To determine rho, the signal from the Test Set’s CDMA generator is
compared with a known-calibrated vector signal analyzer. This is done
by setting the Test Set’s CDMA generator in forward channel (QPSK)
mode and then measuring modulation accuracy with the vector signal
analyzer. Since the vector signal analyzer cannot measure rho directly,
the EVM (Error Vector Magnitude) % rms is measured and rho is
calculated from the EVM data.
Setup
Figure 8-34 CDMA Analyzer Modulation Accuracy Test 29
Procedure
1. On the Test Set:
a. Press the
Preset key.
b. Press the
CDMA Gen key. The CDMA GENERATOR screen
appears, see Figure 8-35.
c. Set the RF Gen Freq to 881.52 MHz.
RF IN/OUT
Vector Signal Analyzer
INPUT
DUPLEX OUT