74 Chapter 3
Troubleshooting
Frequently Encountered Diagnostic Messages
Time-outs
Certain failures may cause a frequency or voltage reading to time out,
that is, the time required for the measurement will be unreasonably
long. If a timeout occurs, measurement execution will stop and an error
message will be displayed.
If frequency or voltage readings have been successfully made before
the timeout, the assembly currently being tested or a multiplexer on
the A2A33 Measurement assembly may be at fault.
If most measurements fail, the A2A23 Reference assembly may be
supplying faulty clock signals to the A2A33 Measurement assembly.
Re-run the test to see if the timeout is intermittent.