194 Chapter 8
Performance Tests
RF Generator Level Accuracy Performance Test 5
+ RF level measured at -80 dBm in step 8i
– RF level measured at -80 dBm in step 8j
+ RF level measured in step 8k
For example, if:
RF power measured at
10 dBm in step 3b or 5b = 10.2 dBm
RF level measured at
80 dBm in step 8i = 70.1 dB
RF level measured at -80 dBm in step 8j =
52.6 dB
RF level measured at -100 dBm in step 8k =
73.2 dB the
corrected level at
100 dBm is 10.2 + (70.1) (52.6) + (73.2) =
100.9 dBm. Compare the corrected values with the limits in the
PTR.