Chapter 2 43
Product Information
Upgrades
14.Enter the firmware code word: This code word is unit specific and
can only be used with the unit for which it was ordered. If you have
more than one test set to upgrade you must associate the upgrade kit
with the serial number of the test set for which it was ordered.
15.Once the codeword is entered the program will prompt you to cycle
power to enable the CDMA 2000 feature.
After the Test set has fully booted you must choose the IS-2000 function
by selecting the Inst Config screen:
Move the cursor to the CDMA Std field on the Instrument Configure
screen and select IS-2000 to enable the CDMA 2000 functionality.
If you encounter any difficulty with this upgrade please email or call
Technical Support for additional assistance.
Checking Firmware Version
The following procedure checks the current version of firmware in the
Test Set.
1. Power on the Test Set.
2. Press the
Inst Config key. The INSTRUMENT CONFIGURE screen
appears. The host firmware level appears in the upper right corner
of this screen.
3. To find the Receive (RX) DSP firmware revision level, select the title
bar of the INSTRUMENT CONFIGURE screen. A drop down menu
appears.
4. Select SERVICE. The SERVICE screen appears, see Figure 2-3 on
page 43.