1100 Series Capillary LC System Manual 51
Capillaries and Fittings 3
Figure 3 Micro column switching valve connections
1
2
3
4
To det e c t o r
From injection valve
Column 1
Column 2
5
6
Tabl e 13 Additional capillaries
Description Fitting
type
Diameter
(µm)
Length
(mm)
Material Volume
(µl)
Pressure
drop (Bar)
Part number
OQ/PV capillary C/D 50 400 PFS
*
0.8 4.4 G1375-87314
MS capillary C/2xD 50 1100 PFS * 2.2 12 5065-9906
MS capillary E/2xD 50 700 PFS * 1.4 7.6 G1375-87313
* PEEK coated fused silica
NOTE
The pressure drops in table 13 are calculated for water (viscosity 1) and for a flow rate of
10 µl/min.
NOTE
The pressure drops in tables 7 to 13 are indicated values at a specific flow rate and with
water (viscosity = 1). For other solvents or other flow rates, use the indicated relation to
calculate the approximate pressure drop. Depending on tolerance of the capillary diameter
the pressure drop values can vary by +/- 25% compared to the calculated results.
Pressure (Bar)=
Flow(µl/min) xViscosity (mPa
*
s) xLength(mm) x 21333 / 3.14 x Diameter4 m)