1100 Series Capillary LC System Manual 141
Options 6
12 Replace the loop capillary with the capillary G1329-87302 if you have a
micro autosampler (G1389A) or with the capillary G1377-87300 if you have
a micro well-plate sampler (G1377/78A).
13 Replace the capillary between the injection valve (port 6) and the column
with the capillary G1375-87312. If a thermostat (G1330A/B) is in place use
the capillary G1375-87306.
14 Replace the capillary between the column and the detector with the
capillary G1315-87311.
The pressures in Table 82 and Table 83 are indicated values, measured on one
system. These values can differ from one system to another.
Don‘t forget to change the loop or syringe size to 40 µl in the injector configuration
windows of the user interface.
Needle seat assembly must be G1329-87101 with the 100 µm capillary (G1389A).
Needle seat assembly must be G1375-87317 with the 100 µm capillary ((G1377/78A).
Above a flow of 200 µl/min it is recommend to rout the flow through the Peltier. The
capillary G1316-87300 is then connected between the Peltier “out” and the column inlet.
Replace the 500 nl flow cell with the standard flow cell (G1315-60012), the semi micro flow
cell (G1315-6001) or the high pressure flow cell (G1315-60015)