1100 Series Capillary LC System Manual 139
Options 6
0.1 - 2.5 ml/min Flow Capillary Kit (5065-4495)
It is possible to use the capillary pump with a flow rate higher than
100 µl/min. For this the pump must be used in the normal mode and some
hardware modifications are necessary.
From 100 to 200 µl/min, bypassing the electronic flow control is needed. No
other hardware modifications are necessary.
From 200 to 2500 µl/min bypassing the electronic flow sensor, installing the
manual purge valve (supplied with the accessory kit), changing the UV
detector cell and changing the capillaries in the flow path is needed.
The 0.1 to 2.5 ml/min flow capillary kit (5065-4495) include all the capillaries
needed to work with a flow from 200 to 2500 µl/min.
* SST: stainless steel
* * PFS: Peek coated fused silica
Tabl e 81 0.1 - 2.5 ml/min flow capillary kit 5065-4495 content
Part number Diameter
(µm)
Pressure
drop (Bar)
Length
(mm)
Material Volume
(µl)
Fitting
type
G1375-87400 170 2 280 SST * 6.4 A/A
G1375-87318 125 15 550 PFS ** 6.8 B/C
G1375-87312 100 13 200 PFS 1.6 B/C
G1329-87302 250 3 1800 SST 88 B/B
G1375-87312 100 13 200 PFS 1.6 B/C
G1375-87306 100 37 550 PFS 4.4 C/B
G1316-87300 170 <1 70 SST 1.6 A/A