1100 Series Capillary LC System Manual 125
Parts and Materials 5
Tabl e 66 Agilent 1100 to 3390 integrators
Connector
01046-60203
Pin
3390
Pin
Agilent 1100
Signal Name Active
(TTL)
2 1 - White Digital ground
NC 2 - Brown Prepare run Low
7 3 - Gray Start Low
NC 4 - Blue Shut down Low
NC 5 - Pink Not connected
NC 6 - Yellow Power on High
NC 7 - Red Ready High
NC 8 - Green Stop Low
NC 9 - Black Start request Low
Tabl e 67 Agilent 1100 to 3392/3 integrators
Connector
01046-60206
Pin
3392/3
Pin
Agilent 1100
Signal Name Active
(TTL)
3 1 - White Digital ground
NC 2 - Brown Prepare run Low
11 3 - Gray Start Low
NC 4 - Blue Shut down Low
NC 5 - Pink Not connected
NC 6 - Yellow Power on High
9 7 - Red Ready High
1 8 - Green Stop Low
NC 9 - Black Start request Low
4 - Key