1100 Series Capillary LC System Manual 163
Options 6
9 Insert the flow cell capillaries and tighten them
fingertight.Use of the wrench and torque adapter as
described page 155 and tighten the fittings alternate
10 Reassemble the flow cell body to the holder.
11 Re-install the flow cell and connect the capillaries to
the union holder.
Remove the flow cell and perform a leak test.
If no leak is observed, install the flow cell and you are
ready to work.
IN
OUT