14 1100 Series Capillary LC System Manual
1 Installing your Capillary LC System
Installing a Capillary LC System with a Thermostatted Micro Sampler
These instructions will produce two stacks of modules. The left-hand stack
will consist of the capillary pump, degasser and solvent cabinet. The
right-hand stack will consist of the thermostatted micro sampler (bottom),
thermostatted column compartment (TCC), and diode array detector (DAD)
on top.
Required cables, tubes and capillaries for each module are included with the
system shipment, or found in the module accessory kits.
Refer to Chapter 3 for detailed information on system plumbing connections,
and Agilent part numbers and descriptions for capillaries throughout the
system.
This manual provides an overview of the entire Capillary LC System. For more
detailed information about each module, refer to the reference manual
provided with each module.
The Agilent part numbers for capillaries indicated in the text are intended for use with the
standard capillary pump, as the pump is shipped from the factory. If the optional Extended
Flow Range Kit (G1376-68707) is to be installed in the pump, several of these capillaries,
throughout the entire system, will be changed. Refer to chapter 6 for detailed information
on the Extended Flow Range Kit.