16 1100 Series Capillary LC System Manual
1 Installing your Capillary LC System
Install the Micro Sampler (G1387A Micro Autosampler (ALS), or G1378A
Micro Well-plate Sampler)
1 Ensure the line power switch at the front of the micro sampler is off.
2 Place the micro sampler on top of the thermostat module. Make sure that
the two modules are interlocked correctly.
3 Remove the sampler shipping protection foam.
4 Connect the free end of the thermostat-to-sampler cable (G1330-81600) to
the 26-pin connector at the rear of the micro sampler.
5 Connect the CAN-bus cable (5181-1519) to one of the CAN connectors at the
rear of the micro sampler.
6 Connect the power cable to the power socket at the rear of the micro
sampler. Connect the power cable to the power socket at the rear of the
thermostat module.
Do not connect these power cables to power until you have finished the
hardware installation of all modules in the stack.
7 Install the air channel adapter (G1329-43200) between the micro sampler
and the thermostat module. See the sampler reference manual for more
detail if required.
8 Connect one end of the sampler-to-column capillary (G1375-87304) to port
6 of the sampler injection valve. The other end of this capillary will later be
connected to the inlet of the analytical column in the TCC.
9 Connect the large-bore corrugated plastic leak drain tubing to the micro
sampler leak drain fitting. Route the leak drain tubing to an appropriate
waste container.
Connect the power cable to the thermostat module power socket only after you have
connected the thermostat-to-sampler cable (G1330-81600) between the thermostat
module and the sampler. Failure to follow this warning will result in damage to the
electronics of the thermostat module and the sampler.
Carefully route all capillaries so that they are not crushed or broken by module front covers.
Avoid excessive bending. See Chapter 2 for advice on handling capillaries.