8 1100 Series Capillary LC System Manual
1 Installing your Capillary LC System
Install the Diode Array Detector (DAD) (G1315B)
Ensure the line power switch at the front of the DAD is off.
1 Ensure the line power switch at the front of the Diode Array Detector
(DAD) is off.
2 If the system is to be connected to the user interface by LAN, install the
JetDirect card into the DAD. See Replacing the Interface Board in the DAD
Reference Manual.
3 Place the DAD on the bench.
4 Connect one end of the LAN cross over cable (5183-4649) to the connector
on the JetDirect card. Connect the other end of the LAN cross over cable to
the Chemstation.
5 Connect the Can-bus cable (5181-1516) to one of the CAN connectors at the
rear of the DAD.
6 Connect the power cable to the power socket at the rear of the DAD. Do not
connect the power cable to power until you have finished the hardware
installation of all modules in the stack.
7 Install the DAD flow cell (G1314-68714).
8 Route the DAD flow cell outlet capillary (G1315-68708) to an appropriate
waste container. The DAD flow cell inlet capillary (G1315-68703) will later
be connected to the outlet of the analytical column.
9 Connect the large-bore corrugated plastic leak drain tubing to the DAD leak
drain fitting. Route the leak drain tubing to an appropriate waste container.
WARNING
Connect the power cable to power only after you have finished the hardware
installation of all modules in the stack.