1100 Series Capillary LC System Manual 7
Installing your Capillary LC System 1
Installing a Capillary LC System with a Non-Thermostatted Sampler
These instructions will produce a single stack of modules, with the Diode
Array Detector (DAD) on the bottom. Required cables, tubes and capillaries
for each module are included with the system shipment, or found in the
module accessory kits.
Refer to Chapter 3, “Capillaries and Fittings,” starting on page 45 for detailed
information on system plumbing connections, and Agilent part numbers and
descriptions for capillaries throughout the system.
This manual provides an overview of the entire Capillary LC System. For more
detailed information about each module, refer to the reference manual
provided with each module.
The Agilent part numbers for capillaries indicated in the text are intended for use with the
standard capillary pump, as the pump is shipped from the factory. If the optional Extended
Flow Range Kit (G1376-68707) is to be installed in the pump, several of these capillaries,
throughout the entire system, will be changed. Refer to chapter 6 for detailed information
on the Extended Flow Range Kit.