12 1100 Series Capillary LC System Manual
1 Installing your Capillary LC System
Install the Micro Vacuum Degasser (G1379A)
1 Ensure the line power switch at the front of the micro vacuum degasser is
off.
2 Place the degasser on top of the pump. Make sure that the two modules are
interlocked correctly.
3 Connect one end of the remote cable (5061-3378) to the rear of the
degasser. Connect the other end of the cable to the remote port at the rear
of the pump.
4 The degasser accessory kit has a set of 4 solvent tubes (G1322-67300). Each
tube is labeled A, B, C or D. Connect each solvent tube to its intended
OUTLET channel port on the degasser.
5 Connect the other end of the solvent tube to its intended port at the pump
solvent selection valve. Follow the guide below:
WARNING
Connect the power cable to power only after you have finished the hardware
installation of all modules in the stack.
Degasser
OUTLET
Pump Solvent Selection
Valve Port
A to A1 (left half, upper)
B to A2 (left half, lower)
C to B1 (right half, upper)
D to B2 (right half, lower)