10 1100 Series Capillary LC System Manual
1 Installing your Capillary LC System
Install the Micro Well-plate Sampler (G1377A)
1 Ensure the line power switch at the front of the micro well-plate sampler
(micro WPS) is off.
2 Place the micro WPS on top of the TCC. Make sure that the two modules are
interlocked correctly.
3 Remove the sampler shipping protection foam.
4 Connect the CAN-bus cable (5181-1519) to one of the CAN connectors at the
rear of the micro WPS.
5 Connect the power cable to the power socket at the rear of the micro WPS.
Do not connect the power cable to power until you have finished the
hardware installation of all modules in the stack.
6 Connect the free end of the Can-bus cable from the TCC to the unused
Can-bus connector at the rear of the micro sampler.
7 Connect one end of the sampler-to-column capillary (G1375-87304) to port
6 of the sampler injection valve. Connect the other end of this capillary to
the inlet of the analytical column in the TCC.
Connect the power cable to power only after you have finished the hardware
installation of all modules in the stack.
Carefully route all capillaries so that they are not crushed or broken by module front covers.
Avoid excessive bending. refer to Chapter 2 for advice on handling capillaries.