1100 Series Capillary LC System Manual 25
Installing your Capillary LC System 1
Conditioning the System Under Method Conditions
If you wish to condition the analytical column at this time, leave the column
installed in the TCC.
If you do not wish to condition the analytical column at this time, then remove
the column. In the TCC, connect the sampler-to-column capillary
(G1375-87304) directly to the DAD flow cell inlet capillary (G1315-68703).
This connection can be made with a ZDV fitting (0100-0900)
Enter your method conditions, and turn on the pump. Allow the system to
equilibrate under these conditions.
Tabl e 3 Choice of Priming Solvents for Different Purposes
Activity Solvent Comments
After an installation Isopropanol Best solvent to flush air out of the
system
After an installation (second
choice)
Ethanol or Methanol Alternative to Isopropanol if no
Isopropanol is available
When switching between reverse
phase and normal phase (both
times)
Isopropanol Best solvent to flush air out of the
system
To clean the system when using
buffers
Bidistilled water Best solvent to re-dissolve salts
After a solvent change Bidistilled water Best solvent to re-dissolve salts
After the installation of normal
phase seals
(P/N 0905-1420)
Hexane +
5% Isopropanol
Good wetting properties
To clean the capillaries Acetone Best solvent to remove impurities
from the capillaries