1100 Series Capillary LC System Manual 169
Performance Specifications 7
Performance Specifications Agilent 1100 Series Thermostatted Micro
Autosampler
Tabl e 89 Performance Specifications Agilent 1100 Series Thermostatted Micro
Autosampler
Type Specification
Sample capacity 100 x 2 ml vials in 1 tray. Microvials (100 or 300 µl) with sleeves
(reduced cooling performance with microvials)
Settable injection volume 0.01 to 8 µl with small loop capillary
0.01 to 40 µl with extended loop capillary
Precision Typically < 0.5 % RSD from 5 – 40 µl,
Typically < 1 % RSD from 1 – 5 µl
Typically < 3 % RSD from 0.2 – 1 µl
Minimum sample volume 1 µl from 5 µl sample in 100 µl microvial, or 1 µl from 10 µl sample
in 300 µl microvial
Carryover Typically < 0.1 % without automated needle wash.
Typically < 0.05 % with external needle cleaning and 1ul injection
volume
Sample viscosity range 0.2 – 5 cp
Recommended pH-range 1.0 – 8.5, solvents with pH < 2.3 should not contain acids which
attack stainless steel. Upper pH range is limited by fused silica
capillaries.
Material in contact with
solvent
Stainless steel, sapphire, PTFE, PEEK, fused silica, Vespel
GLP features Early maintenance feedback (EMF), electronic records of
maintenance and errors
Communications Controller-area network (CAN). GPIB (IEEE-448), RS232C,
APG-remote standard, optional four external contact closures and
BCD vial number output
Safety features Leak detection and safe leak handling, low voltages in
maintenance areas, error detection and display
Housing All material recyclable