142 1100 Series Capillary LC System Manual
6Options
.
.
Tabl e 82 Pressure drop at 2.5 ml/min for different concentrations (no column)
% of organic phase Pressure (bar) for Methanol Pressure (bar) for Acetonitrile
0165162
20 170 169
40 158 154
60 132 128
80 100 95
100 75 72
Tabl e 83 Pressure drop for different columns and different flow rates, with a gradient
from 0 to 100% Acetonitrile in 10 minutes.
Column (id and length) Flow rate (ml/min) Pressure (bar)
100 x 2.1 mm 0.4 92 (max.) 38 (lowest)
100 x 2.1 mm 0.8 174 (max.) 68 (lowest)
125 x 4.0 mm 1.0 131 (max.) 45(lowest)
125 x 4.0 mm 1.5 190 (max.) 67 (lowest)
100 x 4.6 mm 2.0 213 (max.) 86 (lowest)
100 x 4.6 mm 2.5 272 (max.) 112 (lowest)