1100 Series Capillary LC System Manual 31
Optimizing Performance 2
To achieve the best flow stability, especially in the micro mode, %Ripple must
be within acceptable values, typically no worse than 2%.
Fused Silica Capillary issues
When you connect a capillary (especially at the column) press it smoothly
into the fitting to avoid air gaps. Incorrect setting will result in dispersion
causing tailing or footing peaks.
Be careful when you bend a Fused Silica Capillary. The diameter must not
be smaller than 40 mm.
When you replace a part, especially a capillary, clean it with Acetone.
If a fused silica capillary leaks, do not retighten under flow. Set column
flow to zero, reinsert the capillary, tighten and set new column flow.
Avoid the use of alkaline solutions (pH > 8.5) which can attack the fused
silica from the capillaries.
Be careful not to crush capillaries when applying module doors.
A broken capillary can release silica particles into the system (e.g. cell)
causing problems in the system down-stream of the break.
A blocked capillary can often be unblocked by back-flushing the capillary.
Acetone is a recommended solvent for capillary back-flushing.
NOTE
Do not overtighten the fused silica capillaries. see Chapter 3, “Capillaries and Fittings,”
starting on page 45 for information on installing and handling capillaries.