Command Reference
Pulse Modulation Subsystem ([:SOURce])
Chapter 4 289
:PULM:INTernal[1]:DELay:STEP
Supported PSG-A Series
[:SOURce]:PULM:INTernal[1]:DELay:STEP <num>[<time suffix>]
[:SOURce]:PULM:INTernal[1]:DELay:STEP?
This command sets the step increment for the pulse delay.
The optional variable [<time suffix>] accepts nS (nano-seconds) to S (seconds).
*RST N/A
Range 10nS to (pulse period
20 nS)
Key Entry N/A
Remarks The value set by this command is used with the UP and DOWN choices
for the pulse modulation delay command. Refer to
“:PULM:INTernal[1]:DELay” on page 288 for more information.
The setting enabled by this command is not affected by signal
generator power-on, preset, or *RST.
:PULM:INTernal[1]:FREQuency
Supported PSG-A Series
[:SOURce]:PULM:INTernal[1]:FREQuency <val><unit>
[:SOURce]:PULM:INTernal[1]:FREQuency?
This command sets the rate of the internal square wave pulse modulation source.
*RST
+
4.00000000E
+
002
Range 0.1HZ–10MHZ
Key Entry
Pulse Rate
Remarks This command is used when SQUare is the current pulse modulation
type. Refer to “:PULM:SOURce:INTernal” on page 292 for the pulse
modulation type selection.