Command Reference
IEEE 488.2 Common Commands
Chapter 4 159
*SAV
Supported All
*SAV <reg>,<seq>
The Save (SAV) command saves the state of the signal generator to the specified memory
register <reg> of the specified sequence <seq>.
*RST N/A
Remarks N/A
*SRE
Supported All
*SRE <data>
The Service Request Enable (SRE) command sets the value of the Service Request
Enable Register.
The variable <val> is the decimal sum of the bits that will be enabled. Bit 6 (value 64) is
ignored and cannot be set by this command.
*RST N/A
Range 0–255
Key Entry N/A
Remarks Refer to “Generating a Service Request” on page 107, “Status Byte
Group” on page 110, and “Service Request Enable Register” on
page 112 for more information.
Entering values from 64 to 127 is equivalent to entering values from 0
to 63.
The setting enabled by this command is not affected by signal
generator preset or *RST. However, cycling the signal generator power
will reset it to zero.
Range Registers: 0–99 Sequences: 0–9
Key Entry
Save Reg Save Seq[n] Reg[nn]