Command Reference
IEEE 488.2 Common Commands
Chapter 4154
IEEE 488.2 Common Commands
*CLS
Supported All
*CLS
The Clear Status (CLS) command clears the Status Byte Register, the Data
Questionable Event Register, the Standard Event Status Register, the Standard
Operation Status Register and any other registers that are summarized in the status
byte.
*RST N/A
Range N/A
Key Entry N/A
Remarks N/A
*ESE
Supported All
*ESE <data>
The Standard Event Status Enable (ESE) command sets the Standard Event Status
Enable Register.
The variable <data> represents the sum of the bits that will be enabled.
*RST N/A
Range 0–255
Key Entry N/A
Remarks The setting enabled by this command is not affected by signal
generator preset or *RST. However, cycling the signal generator power
will reset this register to zero.
Refer to “Standard Event Status Group” on page 114 and “Standard
Event Status Enable Register” on page 116 for more information.