Command Reference
Correction Subsystem ([:SOURce]:CORRection)
Chapter 4234
:FLATness:PAIR
Supported All
[:SOURce]:CORRection:FLATness:
PAIR <freq>[<freq suffix>],<corr>[<corr suffix>]
This command sets a frequency and amplitude correction pair.
*RST N/A
Key Entry
Configure Cal Array
Remarks The maximum number of points that can be entered is 1601.
:FLATness:POINts?
Supported All
[:SOURce]:CORRection:FLATness:POINts?
This command queries the signal generator for the number of points in the user flatness
correction file.
*RST N/A
Range N/A
Key Entry N/A
Remarks N/A
Range 20 GHz Models
Frequency: 100kHZ–20GHZ Correction (Std.):
20 to 25DB
Correction (Opt. 1E1):
135 to 25DB
40 GHz Models
Frequency: 100kHZ–40GHZ Correction (Std.):
20 to 25DB
Correction (Opt. 1E1):
135 to 25DB