Command Reference
List/Sweep subsystem ([:SOURce])
Chapter 4256
:LIST:DWELl:TYPE
Supported All
[:SOURce]:LIST:DWELl:TYPE LIST|STEP
[:SOURce]:LIST:DWELl:TYPE?
This command toggles the dwell time for the list sweep points between the values
defined in the list sweep and the value for the step sweep.
LIST This choice selects the dwell times from the list sweep. Refer to
“:LIST:DWELl” on page 255 for setting the list dwell points.
STEP This choice selects the dwell time from the step sweep. Refer to
“:SWEep:DWELl” on page 262 for setting the step dwell.
*RST LIST
Key Entry
Dwell Type List Step
Remarks N/A
:LIST:FREQuency
Supported All
[:SOURce]:LIST:FREQuency <val>{,<val>}
[:SOURce]:LIST:FREQuency?
This command sets the frequency values for the current list sweep points.
The variable <val> is measured in units of hertz.
*RST N/A
Key Entry N/A
Remarks The setting enabled by this command is not affected by signal
generator power-on, preset, or *RST.
Choices LIST STEP
Range 20 GHz Models: 100E3
−
20E9 40 GHz Models: 100E3
−
40E9