Agilent Technologies PSG Manual de Instruções Página 299

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 362
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 298
Command Reference
Power Subsystem ([:SOURce])
Chapter 4 285
:POWer:STARt
Supported All
[:SOURce]:POWer:STARt <val><unit>
[:SOURce]:POWer:STARt?
This command sets the amplitude of the first point in a step sweep.
*RST
1.35000000E
+
002
Range Refer to “:POWer[:LEVel][:IMMediate][:AMPLitude]” on page 287 for
output power ranges.
Key Entry
Ampl Start
Remarks During an amplitude sweep operation, signal generators with Option
1E1 protect the step attenuator by automatically switching to
attenuator hold (OFF) mode. The amplitude sweep range is limited to
45 dB and be moved around the whole power range by inputting a
different power level.
:POWer:STOP
Supported All
[:SOURce]:POWer:STOP <val><unit>
[:SOURce]:POWer:STOP?
This command sets the amplitude of the last point in a step sweep.
*RST
1.35000000E
+
002
Range Refer to “:POWer[:LEVel][:IMMediate][:AMPLitude]” on page 287 for
output power ranges.
Key Entry
Ampl Stop
Remarks During an amplitude sweep operation, signal generators with Option
1E1 protect the step attenuator by automatically switching to
attenuator hold (OFF) mode. The amplitude sweep range is limited to
45 dB and be moved around the whole power range by inputting a
different power level.
Vista de página 298
1 2 ... 294 295 296 297 298 299 300 301 302 303 304 ... 361 362

Comentários a estes Manuais

Sem comentários