Command Reference
Status Subsystem (:STATus)
Chapter 4 205
:QUEStionable:POWer:PTRansition
Supported All
:STATus:QUEStionable:POWer:PTRansition <value>
:STATus:QUEStionable:POWer:PTRansition?
This command determines what bits in the Data Questionable Power Condition Register
will set the corresponding bit in the Data Questionable Power Event Register when that
bit has a positive transition (0 to 1).
The variable <value> is the sum of the decimal values of the bits that you want to
enable.
*RST N/A
Range 0–32767
Key Entry N/A
Remarks Refer to “Data Questionable Power Status Group” on page 124 for more
information.