Command Reference
Communication Subsystem (:SYSTem:COMMunicate)
Chapter 4 165
:PMETer:CHANnel
Supported All
:SYSTem:COMMunicate:PMETer:CHANnel A|B
:SYSTem:COMMunicate:PMETer:CHANnel?
This command sets the measurement channel on the power meter that is controlled by
the signal generator.
*RST N/A
Key Entry
Meter Channel A B
Remarks A single-channel power meter uses channel A and selecting channel B
will have no effect.
The setting enabled by this command is not affected by signal
generator power-on, preset, or *RST.
The power meter is controlled only through a GPIB cable.
:PMETer:IDN
Supported All
:SYSTem:COMMunicate:PMETer:IDN E4418B|E4419B|E4416A|E4417A
:SYSTem:COMMunicate:PMETer:IDN?
This command sets the model number of the power meter that is controlled by the signal
generator.
*RST N/A
Key Entry
Power Meter
Remarks The setting enabled by this command is not affected by signal
generator power-on, preset, or *RST.
The power meter is controlled only through a GPIB cable.