Command Reference
System Subsystem (:SYSTem)
Chapter 4208
System Subsystem (:SYSTem)
:CAPability
Supported All
:SYSTem:CAPability?
This command queries the signal generator’s capabilities and outputs the appropriate
specifiers:
(RFSOURCE WITH((AM|FM|PULM|PM|LFO)&(FSSWEEP|FLIST)&(PSSWEEP|PLIST)
&TRIGGER&REFERENCE))
This is a list of the SCPI-defined basic functionality of the signal generator and the
additional capabilities it has in parallel (a&b) and singularly (a|b).
*RST N/A
Range N/A
Key Entry N/A
Remarks N/A
:ERRor[:NEXT]
Supported All
:SYSTem:ERRor[:NEXT]?
This command queries the signal generator’s error queue and displays the error message
when available. If there are no error messages, the query returns the following output:
+0,"No error"
When there is more than one error message, the query will need to be sent for each
message.
*RST N/A
Range N/A
Remarks The error messages are erased after being queried.