Agilent Technologies PSG Manual de Instruções Página 272

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 362
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 271
Command Reference
List/Sweep subsystem ([:SOURce])
Chapter 4258
:LIST:MODE
Supported All
[:SOURce]:LIST:MODE AUTO|MANual
[:SOURce]:LIST:MODE?
This command sets the operating mode for the current list or step sweep.
AUTO This choice enables the selected sweep type to perform a sweep of all
points.
MANual This choice enables you to select a sweep point which controls the
frequency and/or amplitude according to the sweep type. Refer to
“:LIST:MANual” on page 257 for selecting a sweep point
*RST AUTO
Key Entry
Manual Mode Off On
Remarks N/A
:LIST:POWer
Supported All
[:SOURce]:LIST:POWer <val>{,<val>}
[:SOURce]:LIST:POWer?
This command sets the amplitude for the current list sweep points.
*RST
Ν/Α
Range Refer to “:POWer[:LEVel][:IMMediate][:AMPLitude]” on page 287 for
output power ranges.
Key Entry N/A
Remarks The setting enabled by this command is not affected by signal
generator power-on, preset, or *RST.
During an amplitude sweep operation, signal generators with Option
1E1 protect the step attenuator by automatically switching to
attenuator hold (OFF) mode. The amplitude sweep range is limited to
45 dB. The 45 dB sweep range can be moved by inputting different
power levels.
Choices AUTO MANual
Vista de página 271