Vista de página 780
Status Reporting 9
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 781
Overload Event Register (:OVLRegister)
The Overload Event Register register hosts these bits:
Name Location Description
Channel 1 OVL bit 0 Overload has occurred on Channel 1 input.
Channel 2 OVL bit 1 Overload has occurred on Channel 2 input.
Channel 3 OVL bit 2 Overload has occurred on Channel 3 input.
Channel 4 OVL bit 3 Overload has occurred on Channel 4 input.
External Trigger
OVL
bit 4 Overload has occurred on External Trigger input.
Channel 1 Fault bit 6 Fault has occurred on Channel 1 input.
Channel 2 Fault bit 7 Fault has occurred on Channel 2 input.
Channel 3 Fault bit 8 Fault has occurred on Channel 3 input.
Channel 4 Fault bit 9 Fault has occurred on Channel 4 input.
External Trigger
Fault
bit 10 Fault has occurred on External Trigger input.