388 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:MTESt:OUTPut
(see page 798)
Command Syntax
:MTESt:OUTPut <signal>
<signal> ::= {FAIL | PASS}
The :MTESt:OUTPut command selects the mask test output condition:
FAIL — the output occurs when there are mask test failures.
PASS — the output occurs when the mask test passes.
You can place the mask test signal on the rear panel TRIG OUT BNC using
the ":CALibrate:OUTPut" on page 223 command.
Query Syntax
:MTESt:OUTPut?
The :MTESt:OUTPut? query returns the currently set output signal.
Return Format
<signal><NL>
<signal> ::= {FAIL | PASS}
See Also "Introduction to :MTESt Commands" on page 372