Agilent Technologies InfiniiVision 6000 Series Manual de Serviço Página 145

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 934
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 144
Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 145
Query Syntax *SRE?
The *SRE? query returns the current value of the Service Request Enable
Register.
Return Format
<mask><NL>
<mask> ::= sum of all bits that are set, 0,..,255;
an integer in NR1 format
See Also "Introduction to Common (*) Commands" on page 128
"*STB (Read Status Byte)" on page 146
"*CLS (Clear Status)" on page 129
5 ESB Event Status Bit Interrupts when enabled conditions in the
Standard Event Status Register (ESR) occur.
4 MAV Message Available Interrupts when messages are in the Output
Queue.
3 --- --- (Not used.)
2 MSG Message Interrupts when an advisory has been displayed
on the oscilloscope.
1 USR User Event Interrupts when enabled user event conditions
occur.
0 TRG Trigger Interrupts when a trigger occurs.
Table 43 Service Request Enable Register (SRE) (continued)
Bit Name Description When Set (1 = High = True), Enables:
Vista de página 144
1 2 ... 140 141 142 143 144 145 146 147 148 149 150 ... 933 934

Comentários a estes Manuais

Sem comentários