Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 579
:TRIGger:M1553:AUTosetup
(see page 798)
Command Syntax
:TRIGger:M1553:AUTosetup
The :TRIGger:M1553:AUTosetup command copies the position, volts/div,
and probe attenuation from the upper threshold channel to the lower
threshold channel, sets the upper/lower trigger levels to +/- 500 mV, turns
on serial decode, and sets the trigger mode to M1553.
See Also "Introduction to :TRIGger Commands" on page 479
":TRIGger:MODE" on page 485
":TRIGger:M1553:SOURce:UPPer" on page 583