Agilent Technologies InfiniiVision 6000 Series Manual de Serviço Página 401

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 934
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 400
Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 401
:MTESt:SCALe:Y2
(see page 798)
Command Syntax
:MTESt:SCALe:Y2 <y2_value>
<y2_value> ::= Y2 value in NR3 format
The :MTESt:SCALe:Y2 command defines the Y2 marker in the coordinate
system for mask testing. All Y values of vertices in the coordinate system
are defined with respect to the boundaries defined by SCALe:Y1 and
SCALe:Y2 according to the following equation:
Y = (Y * (Y2 - Y1)) + Y1
Thus, if you set Y1 to 100 mV, and Y2 to 1 V, a Y value of 0.100 in a
vertex is at 190 mV.
The Y2 value is a voltage value specifying the location of the Y2 marker.
Query Syntax
:MTESt:SCALe:Y2?
The :MTESt:SCALe:Y2? query returns the current setting of the Y2 marker.
Return Format
<y2_value><NL>
<y2_value> ::= Y2 value in NR3 format
See Also "Introduction to :MTESt Commands" on page 372
":MTESt:SCALe:BIND" on page 397
":MTESt:SCALe:X1" on page 398
":MTESt:SCALe:XDELta" on page 399
":MTESt:SCALe:Y1" on page 400
Vista de página 400
1 2 ... 396 397 398 399 400 401 402 403 404 405 406 ... 933 934

Comentários a estes Manuais

Sem comentários