Obsolete and Discontinued Commands 7
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 727
:MEASure:TDELta
(see page 798)
Query Syntax
:MEASure:TDELta?
The :MEASure:TDELta? query returns the time difference between the
Tstop marker (X2 cursor) and the Tstart marker (X1 cursor).
Tdelta = Tstop - Tstart
Tstart is the time at the start marker (X1 cursor) and Tstop is the time at
the stop marker (X2 cursor). No measurement is made when the
:MEASure:TDELta? query is received by the oscilloscope. The delta time
value that is output is the current value. This is the same value as the
front-panel cursors delta X value.
Return Format
<value><NL>
<value> ::= time difference between start and stop markers in NR3 format
See Also "Introduction to :MARKer Commands" on page 308
"Introduction to :MEASure Commands" on page 324
":MARKer:X1Position" on page 310
":MARKer:X2Position" on page 312
":MARKer:XDELta" on page 314
":MEASure:TSTArt" on page 731
":MEASure:TSTOp" on page 732
NOTE
The :MEASure:TDELta command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:XDELta command (see page 314) instead.