Agilent Technologies InfiniiVision 6000 Series Manual de Serviço Página 522

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 934
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 521
522 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TRIGger:FLEXray:EVENt:TYPE
(see page 798)
Command Syntax
:TRIGger:FLEXray:EVENt:TYPE <event>
<event> ::= {WAKeup | TSS | {FES | DTS} | BSS}
Selects the FlexRay event to trigger on. The event setting is only valid
when the FlexRay trigger mode is set to EVENt.
WAKeup — triggers on Wake- Up event.
TSS — triggers on Transmission Start Sequence event.
FES — triggers on Frame End Sequence event.
DTS — triggers on Dynamic Trailing Sequence event.
BSS — triggers on Byte Start Sequence event.
Query Syntax
:TRIGger:FLEXray:EVENt:TYPE?
The :TRIGger:FLEXray:EVENt:TYPE? query returns the currently selected
FLEXray event.
Return Format
<event><NL>
<event> ::= {WAK | TSS | {FES | DTS} | BSS}
See Also "Introduction to :TRIGger Commands" on page 479
":TRIGger:FLEXray:TRIGger" on page 528
":TRIGger:FLEXray:AUTosetup" on page 518
":TRIGger:FLEXray:SOURce" on page 527
NOTE
FES and DTS are equivalent.
NOTE
This command is only valid when the FLEXray triggering and serial decode option (Option
FLX) has been licensed.
Vista de página 521
1 2 ... 517 518 519 520 521 522 523 524 525 526 527 ... 933 934

Comentários a estes Manuais

Sem comentários